The SOPs inspectors expect for computerised systems
A set of procedures for computerised systems is one of the first things an inspector asks to see, and one of the easiest to criticise. The usual problem is not a missing document. It is a procedure that says one thing while the records show another, or one that leaves out the detail the regulations actually ask for. This article maps the core set of procedures to the texts behind them, and shows what each should contain to stand up to questions.
Why procedures matter in the regulations
The wording varies, but the expectation is steady. 21 CFR 11.10 says persons who use closed systems to create, modify, maintain or transmit electronic records shall employ procedures and controls designed to ensure the authenticity, integrity and, when appropriate, the confidentiality of electronic records. Among the listed controls are training and experience for the people who use the systems (11.10(i)), written policies that hold individuals accountable for actions under their electronic signatures (11.10(j)), and controls over system documentation, including revision and change control procedures that keep an audit trail of changes (11.10(k)).
Annex 11 (2011) uses the same pattern. Change and configuration management should follow a defined procedure (clause 10), and personnel need appropriate qualifications, access and defined responsibilities (clause 2). Procedures are how those expectations become something you can show.
The core set, and what each should contain
The table below maps ten procedures to the clause or section that drives them. It is our synthesis of the sources, so treat it as a checklist to compare against your own document set, not as a regulatory list.
| Procedure | What it should cover | Source |
|---|---|---|
| User access and administrator control | Who may be granted access, how access is requested, changed and removed, and that these actions are recorded. Administrators independent of the people responsible for record content. Periodic review of accounts. | Annex 11 cl. 12.1 to 12.3; 21 CFR 11.10(d); FDA data integrity Q&A; draft Annex 11 s. 11 |
| Audit trail review | Who reviews, what is reviewed, how often, and what happens when something unexpected is found. The frequency set from process knowledge and risk. | Annex 11 cl. 9; FDA data integrity Q&A, question 7; draft Annex 11 s. 12.5 to 12.8 |
| Backup and restore | What is backed up, how often, where copies are kept, and how restores are tested and recorded. | Annex 11 cl. 7.2; PIC/S PI 041-1; draft Annex 11 s. 16 |
| Change and configuration control | How changes, including configuration changes, are requested, assessed for impact, approved, tested and recorded. | Annex 11 cl. 10; 21 CFR 11.10(k)(2) |
| Periodic review | How systems are reviewed to confirm they remain valid and compliant, and what the review covers. | Annex 11 cl. 11; draft Annex 11 s. 14 |
| Incident and deviation handling | That all incidents, not only failures, are reported and assessed, with root cause and corrective action for critical ones. | Annex 11 cl. 13 |
| Supplier management | How suppliers are selected, assessed, contracted and monitored, with the file kept for inspectors. | Annex 11 cl. 3 and 4.5 |
| Validation and risk assessment | How the extent of validation is decided from a documented risk assessment, and what the validation records must hold. | Annex 11 cl. 1 and 4 |
| Data storage, archiving and retention | How data is secured, checked for readability and accessibility, archived, and retrieved through the retention period. | Annex 11 cl. 7.1 and 17; MHRA data integrity guidance |
| Business continuity | What happens when a critical system is down, how quickly the alternative is brought into use, and how it is tested. | Annex 11 cl. 16 |
What good looks like in three of them
Audit trail review
This is the procedure most often asked about, because it is where data integrity expectations meet daily work. The draft revision of Annex 11 (still a draft, not a requirement) is specific about what a procedure should say: who should make the review, what should be reviewed and when. It also says reviewing every entry may not be effective, so reviews should be targeted by risk and focus on detecting changes to critical processes or data, and that a key element is verifying the reason a change was made (draft sections 12.5 and 12.7). Even under the current text, a procedure that states these points is easier to defend than one that says only "audit trails are reviewed periodically".
User access
FDA's data integrity guidance recommends restricting the ability to alter specifications, process parameters, data, or manufacturing or testing methods by technical means where possible, for example by limiting permissions to change settings or data. It also says firms should document each authorised person's access privileges for every CGMP computer system in use. A procedure should therefore describe roles and permissions, not just a general statement that access is restricted.
Backup and restore
Annex 11 clause 7.2 says the ability to restore data should be checked during validation and monitored periodically. A backup procedure that stops at "a backup is taken nightly" misses half the requirement. It should say how restore is tested, how often, and where the results are recorded.
Why procedures get picked apart
Reading the texts, three weaknesses are easy for an inspector to find. They are inferences from how the rules are written, not statistics about what inspectors find.
- Procedure and practice differ. The regulations expect procedures to be followed as well as written. If an SOP says reviews are monthly and the records show otherwise, the SOP becomes evidence of a gap.
- The detail is missing. General statements such as "access is controlled" or "data is backed up" do not say who, what, when or how. Clauses such as Annex 11 clause 9 and 12.3 are about specifics.
- The procedure ignores the system. One generic SOP applied to a stand-alone instrument, a cloud application and a spreadsheet cannot describe all three accurately. Several sources (Annex 11 clause 12.2, MHRA section 3) tie controls to criticality and risk, so the procedure should too.
FDA's requests in a 2026 warning letter to a drug manufacturer show what regulators look for in the procedures behind a computerised system. They included detailed procedures for robust use and review of audit trail data, the current status of audit trail implementation on each system, and procedural updates with training on system security, user roles and secondary review of analyses. That list mirrors the table above.
Ownership and training
Every procedure needs an owner who is responsible for keeping it accurate. Annex 11 defines the system owner as the person responsible for the availability and maintenance of a computerised system and the security of the data on it, and the process owner as the person responsible for the business process. Clause 2 expects close cooperation between them, with QA, IT and, where relevant, Qualified Persons. Part 11 adds that people who develop, maintain or use these systems must have the education, training and experience for their tasks (11.10(i)). A procedure with no owner, or one that people have never been trained on, tends to drift away from practice, which is the first weakness above.
A quick self-check
For each procedure in your set, ask:
- Does it name who does what, and when?
- Does it match what people actually do, and is there a record that shows it?
- Does it say how the approach changes with the risk or criticality of the system?
- Are the people who follow it trained on the current version (21 CFR 11.10(i))?
- Is it under change control, so a change leaves a trail (21 CFR 11.10(k)(2))?
A procedure that fails any of these is worth fixing before somebody else reads it.
Key takeaways
- Part 11 and Annex 11 both expect procedures and controls, not just good intentions.
- A core set of ten procedures covers access, audit trails, backup and restore, change, periodic review, incidents, suppliers, validation, storage and archiving, and continuity.
- Detail is what makes a procedure defensible: who, what, when, how, and how the answer changes with risk.
- Check that each procedure matches practice and that records prove it.
- Keep procedures under change control and make sure people are trained on the current version.
Further Reading
- 21 CFR Part 11, electronic records and electronic signatures: section 11.10 on controls for closed systems.
- European Commission, EudraLex Volume 4, Annex 11: Computerised Systems (2011): the clauses referred to in the table.
- European Commission, stakeholder consultation on Annex 11 revision: the draft revised Annex 11, still a draft, with more detail on audit trail review, access reviews and backup.
- FDA, Data Integrity and Compliance With Drug CGMP: Questions and Answers: access, administrator roles and audit trail review.
- PIC/S PI 041-1, Good Practices for Data Management and Integrity in Regulated GMP/GDP Environments: storage and backup expectations.
- FDA warning letter 320-26-78 to Laboratorios Jaloma S.A. de C.V., 22 May 2026: an example of the procedure-related information FDA requested.
Want articles like this first?
Members get each new article before it is published here, plus the CSV 101 course, the templates and SOPs library, Regulatory Watch and Ask Iain.
This article is general information for people who work in regulated industries. It is not legal or regulatory advice. Check the regulator's own publications and your own quality system before you act on it.